The SI-111-PW, manufactured by Apogee, is a version of our SI-111 that is fitted with a connector for attachment to a prewired enclosure. This precision infrared radiometer determines the surface temperature of an object without physical contact. The SI-111-PW measures both the subject's surface temperature and the sensor-body temperature. A Campbell Scientific data logger uses these measurements to calculate the correct temperature of the subject.
続きを読む 
				 
							 
							The SI-111-PW consists of a thermopile, which measures surface temperature, and a thermistor, which measures sensor body temperature. The two temperature sensors are housed in a rugged aluminum body that contains a germanium window. 
Both the thermopile and the thermistor output a millivolt signal that most of our data loggers can measure. The data logger uses the Stefan-Boltzman equation to correct for the effect of sensor body temperature on the target temperature. The corrected readings yield an absolute accuracy of ±0.2°C from -10° to +65°C.
The SI-111-PW has a 22-degree half-angle field of view (FOV). The FOV is reported as the half-angle of the apex of the cone formed by the target (cone base) and the detector (cone apex). The target is a circle from which 98% of the radiation viewed by the detector is being emitted.
| Input Power | 2.5 V excitation (for thermistor) | 
| Response Time | < 1 s (to changes in target temperature) | 
| Target Temperature Output Signal | 60 μV per °C difference from sensor body | 
| Body Temperature Output Signal | 0 to 2500 mV | 
| Optics | Germanium lens | 
| Wavelength Range | 8 to 14 μm (corresponds to atmospheric window) | 
| Field of View (FOV) | 22° half angle | 
| Operating Temperature Range | -55° to +80°C | 
| Operating Relative Humidity Range | 0 to 100% RH | 
| Cable Description | 4.5 m (14.76 ft) twisted, shielded 4-conductor wire with Santoprene casing, ending in pigtails | 
| Absolute Accuracy | 
 | 
| Uniformity | 
 | 
| Repeatability | 
 | 
| Diameter | 2.3 cm (0.9 in.) | 
| Length | 6 cm (2.4 in.) | 
| Weight | 190 g (6.7 oz) | 
注意: 以下は代表的な互換性情報を示しています。互換性のある製品や互換性のない製品をすべて網羅したリストではありません。
| 製品 | 互換性 | 注意 | 
|---|---|---|
| 21X (リタイア) | ||
| CR10 (リタイア) | ||
| CR1000 (リタイア) | ||
| CR10X (リタイア) | ||
| CR200X (リタイア) | ||
| CR206X (リタイア) | ||
| CR211X (リタイア) | ||
| CR216X (リタイア) | ||
| CR23X (リタイア) | ||
| CR295X (リタイア) | ||
| CR3000 (リタイア) | ||
| CR500 (リタイア) | ||
| CR5000 (リタイア) | ||
| CR510 (リタイア) | ||
| CR800 (リタイア) | ||
| CR850 (リタイア) | ||
| CR9000 (リタイア) | ||
| CR9000X (リタイア) | 
The SI-111-PW is often fastened to a CM200-series crossarm, a tripod or tower mast, or a user-supplied pole using a CM230, CM230XL, or CM220 mount. The CM230 and CM230XL are adjustable inclination mounts that allow the SI-111-PW to be mounted perpendicular to the target surface when the target surface is on an incline. The CM230XL is similar to the CM230, but the CM230XL places the SI-111-PW further from the pole or crossarm. The SI-111-PW may also be attached directly to a user-supplied camera tripod.
The cable on the SI-111-PW is fitted with a connector that attaches to a prewired enclosure.
SI-111-PWに関するよくある質問の数: 7
すべて展開すべて折りたたむ
There are two accuracy specifications listed for the SI-111:
Using a radiation shield with the sensor helps keep the sensor body temperature in close approximation to the ambient air temperature. Ultimately, the need to protect the sensor from short-wave radiation is dependent on what is being measured and under what conditions. For example, Campbell Scientific recommends using a radiation shield for canopy measurements.
As a general recommendation, recalibration should be done every two years.
The window in the Apogee infrared sensor is inset and protected, but it can become partially blocked in three ways:
Clean the inner threads and sensor window using a cotton swab dipped in the appropriate solvent. For additional cleaning information, see the Maintenance section of the instruction manual.
The SI-111 has been successfully calibrated with a maximum cable length of 100 m without a loss in accuracy. Ideally, the sensor is calibrated with the overall cable length already configured. Our tests have shown, however, that adding a few meters of cable in the field has a negligible effect on the calibration, provided the measurement device has significantly high input impedance, such as with the CR1000.
The SI-111 can be used to measure a wide variety of surfaces, including water and snow. When measuring objects with low emissivity, however, it is particularly important to apply corrections to the measurement.